Reported Failure
Monitor will not read patient's BP but keeps pumping up. I checked it with my simulator and it worked normally but when put on a patient's arm it would not read the BP. I tried a new cuff and hose and got the same result. I think the sensor is going bad. Also, the pump takes an excessive amount of time to pump up. PLEASE make sure to test this out on someone's arm and not just a simulator before sending it back.